Introduction
A dissolved oxygen sensor in water is a device that is used to measure the amount of oxygen that is dissolved in water. It is a critical sensor for monitoring the health and quality of aquatic environments, such as rivers, lakes, and oceans. The sensor works by using a membrane that is sensitive to oxygen, which is exposed to the water.
The oxygen in the water diffuses through the membrane and reacts with the sensor, generating a signal that is proportional to the amount of oxygen present. This information can then be used to assess the health of the water and make informed decisions about managing it.
-
Specification
Product | Dissolved oxygen sensor |
Model | SIN-DO7012 |
Measure range | 0-20mg/L,0-20ppm,0-45deg C |
Accuracy | Resolution: ±3%, Temperature: ±0.5℃ |
Pressure range | ≤0.3 Mpa |
Power supply | 12 VDC |
Main material | SUS316L+PVC (Ordinary Version), |
Titanium Alloy (Seawater Version) | |
O-ring: Fluoro-rubber; Cable: PVC | |
Cable length | Standard 10-Meter Cable, Max: 100m |
Display | 128 * 64 dot matrix LCD with LED backlight |
Power supply | 12 VDC |
Calibration | Automatic air calibration, Sample calibration |
Storage Temperature | -15~65℃ |
-
Communication Protocol
MODBUS-RTU | |
Baud Rate | 4800/9600/19200/38400/57600 |
Data Bits | 8 bit |
Parity Check | no |
Stop Bit | 1bit |
Register Name | Address Location | Data Type | Length | Read/Write | Description |
Dissolved Oxygen Value | 0 | F(Float) | 2 | R(only read) | |
Dissolved Oxygen Concentration | 2 | F | 2 | R | |
Temperature | 4 | F | 2 | R | |
Slope | 6 | F | 2 | W/R | Range:0.5-1.5 |
Deviation Value | 8 | F | 2 | W/R | Range:-20-20 |
Salinity | 10 | F | 2 | W/R | |
Atmospheric Pressure | 12 | F | 2 | W/R | |
Baud Rate | 16 | F | 2 | R | |
Slave Address | 18 | F | 2 | R | Range: 1-254 |
Response Time of Read | 20 | F | 2 | R | |
Modift Baud Rate | 16 | Signed | 1 | W | |
Modify Slave Address | 17 | Signed | 1 | W | Range: 1-254 |
Modify Response Time | 30 | Signed | 1 | W | 6-60s |